  Abstract

A rapid and sensitive chemiluminescence (CL) method using flow-injection (FI) has been developed for the determination of mandelic acid which is based on the enhancement of mandelic acid to the CL intensity of Ru(bipy)32+-Ce(IV) system. The enhancement effect was dependent on the concentration of mandelic acid, based on which, CL system was established for the determination of mandelic acid. The concentrations of Ru(bipy32+, Ce(IV), and 2S4 were optimized. Under the optimum experimental conditions, the linear range and detection limit are 1.46-342.0 μg/mL and 0.072 μg/mL, respectively. The correlation coefficient (R) was 0.99732. The utility of this method was demonstrated by determining mandelic acid in capsules and human urine sample.

  Statistics
Cumulative Counts from November, 2022
Multiple requests among the same browser session are counted as one view. If you mouse over a chart, the values of data points will be shown.


  Cite this article

[IEEE Style]

S. M. Alam, C. W. Jeon, M. M. Karim, S. H. Lee, S. M. Wabaidur, Y. S. Suh, H. Y. Chung, "A Sensitive Procedure for the Rapid Determination of Mandelic Acid by Flow Injection Analysis and Chemiluminescence Detection," Bulletin of the Korean Chemical Society, vol. 30, no. 1, pp. 102-106, 2009. DOI: 10.5012/bkcs.2009.30.1.102.

[ACM Style]

Seikh Mafiz Alam, Chi Wan Jeon, Mohammad Mainul Karim, Sang Hak Lee, Saikh Mohammad Wabaidur, Yeoun Suk Suh, and Hye Young Chung. 2009. A Sensitive Procedure for the Rapid Determination of Mandelic Acid by Flow Injection Analysis and Chemiluminescence Detection. Bulletin of the Korean Chemical Society, 30, 1, (2009), 102-106. DOI: 10.5012/bkcs.2009.30.1.102.